P-DERM® PS-2012 is a trilaminate consisting of high adhesion silicone gel adhesive, polyurethane film and medical pressure sensitive acrylic. PS-2012 is supplied with a 100 micron colorless LDPE liner on the silicone gel adhesive side and a bleached kraft siliconized release liner on the acrylic side of the construction.
P-DERM®
PS-2012
• Excellent Instantaneous Tack
• Atraumatic Removal from Skin
• Hypoallergenic
• Ability to be Removed and Repositioned
• Excellent Converting Properties
• Atraumatic Removal from Skin
• Hypoallergenic
• Ability to be Removed and Repositioned
• Excellent Converting Properties
Property | Value | Test Method |
---|---|---|
Thickness (excluding liner) | 0.231mm | QSP-726 |
Silicone Gel Coat Weight (before perforation) | 150 gsm | QSP-724 |
Carrier Type | Polyurethane Film | -- |
Carrier Thickness | 25 micron | QSP-726 |
Acrylic Adhesive | Henkel 1101 | -- |
Acrylic Coat Weight | 56 gsm | QSP-724 |
Silicone Adhesion | 1.5 N/25mm | QSP-723 |
Silicone Liner Release | 0.053 N/25mm | QSP-723 |
Acrylic Adhesion | 3.3 N/25mm | QSP-722 |
Acrylic Liner Release | 0.176 N/25mm | QSP-722 |
Continuous Use Conditions | 10 - 70 C | QSP-754 |
Biocompatibility | Cytotoxicity: Pass Primary Dermal: Pass Buehler Sensitization: Pass | ISO 10993 |
Recommended Sterilization Method | Ethylene Oxide | -- |
Specific tests should be performed by the end user to determine the product suitability for the particular application.