P-DERM® PS-2041 is a trilaminate consisting of high adhesion silicone gel adhesive, polyurethane film and medical pressure sensitive acrylic. PS-2041 is supplied with a 75 micron polycarbonate liner on the silicone gel adhesive side and an 78# bleached kraft siliconized release liner on the acrylic side of the construction.
P-DERM®
PS-2041
• Excellent Instantaneous Tack
• Atraumatic Removal from Skin
• Hypoallergenic
• Ability to be Removed and Repositioned
• Excellent Converting Properties
• Atraumatic Removal from Skin
• Hypoallergenic
• Ability to be Removed and Repositioned
• Excellent Converting Properties
Property | Value | Test Method |
---|---|---|
Acrylic Adhesion | 4.3 N/25mm | QSP-722 |
Acrylic Coat Weight | 37 gsm | QSP-724 |
Recommended Sterilization Method | Ethylene Oxide | -- |
Continuous Use Conditions | 10 - 70 C | QSP-754 |
Biocompatibility | Cytotoxicity: Pass; Primary Dermal: Pass; Buehler Sensitization: Pass | ISO 10993 |
Silicone Adhesion | 1.6 N/25mm | QSP-723 |
Carrier Thickness | 25 micron | QSP-726 |
Carrier Type | Polyurethane Film | -- |
Silicone Gel Coat Weight | 150 gsm | QSP-724 |
Thickness (excluding liners) | 0.22mm | QSP-726 |
Specific tests should be performed by the end user to determine the product stability for the particular application.