P-DERM® PS-2066 is a trilaminate consisting of a perforated high adhesion silicone gel adhesive, polyurethane film and medical pressure sensitive acrylic adhesive. PS-2066 is supplied with a perforated 75 micron polycarbonate liner on the silicone gel adhesive side and a 30 micron non-perforated siliconized white polypropylene liner on the acrylic side of the construction.

P-DERM
PS-2066
Hypoallergenic
Ability to be Removed and Repositioned
Atraumatic Removal from Skin
Excellent Instantaneous Tack
Excellent Converting Properties
Ability to be Removed and Repositioned
Atraumatic Removal from Skin
Excellent Instantaneous Tack
Excellent Converting Properties
Property | Value | Test Method |
---|---|---|
Perforation Hole Size/Open Area | 1.5mm / 17% | -- |
Acrylic Adhesion | 3.5 N/25mm | QSP-724 |
Acrylic Coat Weight | 37 gsm | QSP-724 |
Recommended Sterilization Method | Ethylene Oxide | -- |
Continuous Use Conditions | 10 - 70 C | QSP-754 |
Biocompatibility | Cytotoxicity: Pass; Primary Dermal: Pass | ISO 10993 |
Silicone Adhesion | 1.2 N/25mm | QSP-723 |
Carrier Thickness | 25 micron | QSP-726 |
Carrier Type | Polyurethane Film | -- |
Silicone Gel Coat Weight | 150 gsm | QSP-724 |
Thickness (excluding liners) | 0.22mm | QSP-726 |