P-DERM® PS-2077 is a 0.31mm thickness silicone gel/acrylic trilaminate consisting of a high adhesion silicone gel skin contact adhesive, polyurethane and medical pressure sensitive acrylic. PS-2077 is supplied with a 75 micron polycarbonate liner on the silicone gel adhesive side and an 84# bleached kraft siliconized liner on the acrylic side of the construction.
P-DERM®
PS-2077
• Excellent Instantaneous Tack
• Atraumatic Removal from Skin
• Hypoallergenic
• Ability to be Removed and Repositioned
• Excellent Converting Properties
• Atraumatic Removal from Skin
• Hypoallergenic
• Ability to be Removed and Repositioned
• Excellent Converting Properties
Property | Value | Test Method |
---|---|---|
Total Thickness (excluding liners) | 0.31mm | QSP-726 |
Silicone Gel Coat Weight | 250 gsm | QSP-724 |
Polyurethane Carrier | 28 micron | QSP-724 |
Acrylic Adhesive Coat Weight | 37 gsm | QSP-724 |
Silicone Gel Adhesion | 1.9 N/25mm | QSP-723 |
Acrylic Adhesion | 4.3 N/25mm | QSP-722 |
Continuous Use Conditions | 10 – 70 C | QSP-754 |
Biocompatibility | Cytotoxicity: Pass Primary Dermal: Pass Buehler Sensitization: Pass | ISO 10993 |
Recommended Sterilization Method | Ethylene Oxide | -- |
Specific tests should be performed by the end user to determine the product suitability for the particular application.